Title of article :
Optimization of Lead-Selective Membrane Electrode Using Central Composite Experimental Design, and Study of Its Behavior with Electrochemical Impedance Spectroscopy
Author/Authors :
مظلوم اردكاني ، محمد نويسنده , , ماندگاري، عظيمه نويسنده Departmentof Chemistry, Payame Noor University, P.O. Box 19395-4697 Tehran, Iran Mandegari, Azimeh , خوشرو، عليرضا نويسنده , , معصوم، سعيد نويسنده , , كارگز، هادي نويسنده Departmentof Chemistry, Payame Noor University, P.O. Box 19395-4697 Tehran, Iran Kargar, Hadi
Issue Information :
دوفصلنامه با شماره پیاپی 0 سال 2016
Pages :
9
From page :
96
To page :
104
Abstract :
در اين كار روش سطح پاسخ همراه با روش طرح تركيب مركزي براي شبيه سازي و بهينه سازي اثر برخي متغيرها (مقادير پلي وينيل كلريد، يونوفر، افزودني و نرم-كننده) بر عملكرد الكترود يون گزين سرب استفاده شده است. شيب 1/29 ميلي ولت در مقادير بهينه از پلي وينيل كلريد (0283/0 گرم)، يونوفر(0074/0 گرم)، افزودني (002/0 گرم) و نرم كننده (060/0 گرم) به دست آمد. محدوده خطي غلظت 1-10×0/1-5-10×0/1 مولار با روش پتانسيومتري براي سرب بدست آمد. تكنيك اسپكتروسكوپي امپدانس الكتروشيميايي براي ارزيابي مكانيسم پاسخ الكترود بكار برده شد. نتايج بدست آمده از اسپكتروسكوپي امپدانس الكتروشيميايي محدوده خطي غلظتي از 6-10× 0/1 تا 1-10×0/1 مولار را نشان مي دهد. همچنين در مقايسه با روش پتانسيومتري محدوده pH بدست آمده از 3 تا 5/5 به مقدار 5/2 تا 0/6 افزايش پيدا كرد.
Abstract :
In this work, response surface methodology in conjunction with central composite design for modeling and optimization of the influence of some process variables (polyvinyl chloride (F1), ionophore (F2), additive (F3) and plasticizer (F4) amounts), on the performance of polyvinyl chloride membrane lead (??) ion-selective electrode is discussed. The slope of 29.1 ± 0.1 mV at the optimal amounts of polyvinyl chloride (0.0283 g), ionophore (0.0074 g), additive (0.002 g) and plasticizer (0.060 g) has been achieved. The electrode exhibited a linear potential response to lead (II) in the concentration range of 1.0 × 10-5 mol L-1 to 1.0 × 10-1 mol L-1 over pH range of 3.0 - 5.5. Greatly, the alternating current impedance technique was applied to investigate the response mechanism of the electrode. The results were obtained from electrochemical impedance spectroscopy shows a linear concentrations range of 1.0 × 10-6 mol L-1 to 1.0×10-1 mol L-1 and in comparison with potentiometry, the pH range increased to 2.5 ? 6.0.
Journal title :
Iranian Journal of Analytical Chemistry
Serial Year :
2016
Journal title :
Iranian Journal of Analytical Chemistry
Record number :
2396114
Link To Document :
بازگشت