Author/Authors :
Tavakoli Golpaygani، A نويسنده Department of Biomedical Engineering, Standard Research Institiute, Karaj, Iran , , Movahedi، M M نويسنده Ionizing and Non-Ionizing Radiation Protection Research Center, Paramedical Sciences School, Shiraz University of Medical Sciences, Shiraz, Iran , , Reza، M نويسنده Department of Electrical Engineering, Iranian Research Organization for Science and Technology (IROST), Tehran, Iran ,
Abstract :
Introduction: Modern medicine employs a wide variety of instruments with
different physiological effects and measurements. Periodic verifcations are routinely
used in legal metrology for industrial measuring instruments. The correct operation of
electrosurgical generators is essential to ensure patient’s safety and management of the
risks associated with the use of high and low frequency electrical currents on human
body.
Material and Methods: The metrological reliability of 20 electrosurgical
equipment in six hospitals (3 private and 3 public) was evaluated in one of the prov
inces of Iran according to international and national standards.
Results: The achieved results show that HF leakage current of ground-referenced
generators are more than isolated generators and the power analysis of only eight units
delivered acceptable output values and the precision in the output power measure
ments was low.
Conclusion: Results indicate a need for new and severe regulations on periodic
performance verifcations and medical equipment quality control program especially
in high risk instruments. It is also necessary to provide training courses for operating
staff in the feld of meterology in medicine to be acquianted with critical parameters to
get accuracy results with operation room equipment.