Title of article :
Method validation of clonidine hydrochloride in human plasma by LC-MS technique
Author/Authors :
Danafar، Hossein نويسنده Zanjan Pharmaceutical Nanotechnology Research Center, School of Pharmacy,Department of Medicinal Chemistry,Zanjan University of Medical Sciences,Zanjan,Iran ,
Issue Information :
فصلنامه با شماره پیاپی سال 2015
Pages :
11
From page :
48
To page :
58
Abstract :
A simple and sensitive high performance liquid chromatography-electrospray ionization mass spectrometry method has been evaluated for the assignment of clonidine hydrochloride in human plasma. The mobile phase composed of acetonitrile–water 60:40 (v/v), and 0.2% formic acid 20 μl of sample was chromatographically analyzed using a repacked ZORBAX-XDB-ODS C18 column (2.1 mm×30 mm, 3.5 micron). Detection of analytes was achieved by tandem mass spectrometry with electrospray ionization (ESI) interface in positive ion mode was operated under the multiple-reaction monitoring mode (m/z 230.0 →213). Sample pretreatment involved in a one-step protein precipitation (PPT) with methanol and percholoric acid (HClO4) of 0.15 mL plasma. Standard curve was linear (r = 0.998) over the concentration range of 0.01-10.0 ng/ml and showed suitable accuracy and precision. The limit of quantification (LOQ) was 0.01ng/ml. The method is rapidly, simple, very steady and precise for the separation, assignment, evaluation of clonidine healthy in human plasma.
Keywords :
Clonidine hydrochloride , LC-MS , Human plasma
Journal title :
Pharmaceutical and Biomedical Research
Serial Year :
2015
Journal title :
Pharmaceutical and Biomedical Research
Record number :
2397043
Link To Document :
بازگشت