Title of article :
Evaluation of the Performance Characteristics of an In-House One Step TaqMan
Real Time-Polymerase Chain Reaction Assay for Detection and Quantification of
Hepatitis C Virus
Author/Authors :
Amini Kafi-abad Sedigheh نويسنده Iranian Blood Transfusion Organization Research Center, Tehran , Ranjbar Kermani Fahimeh نويسنده Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine, Tehran, IR Iran , Samiee Shahram نويسنده Research Center Laboratory, Iranin Blood Transfusion Organization , Sharifi Zohre نويسنده Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine, Tehran, IR Iran , Mousavi Hosseini Kamran نويسنده Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine, Tehran, Iran Mousavi Hosseini Kamran , Maghsoudlu Mahtab نويسنده Department of Research and Education and the, Blood
Transfusion Research Center, High Institute for Research and
Education in Transfusion Medicine, Tehran, IR Iran
Abstract :
Background With the improvement of quantitative molecular
hepatitis C virus (HCV) RNA assays, the usefulness of these assays has
been indicated for management of HCV infection. Recently, various real
time assays with different methodology and performance characteristics
have been introduced. Objectives This study aimed at designing,
developing, and evaluating an in-house 1 step TaqMan real
time-polymerase chain reaction (RT-PCR) assay for detection and
quantification of HCV-RNA. Methods The primers and probe were selected
from a highly conserved region of the HCV genome, which allowed the
detection of 4 common HCV genotypes in Iran. Using 4 quantification
standards from 101 IU/µL to
104 IU/µL and clinical specimens, the current
study determined analytical sensitivity, linear range, precision,
analytical and clinical specificity, and validity of the assay. Data was
analyzed by the SPSS statistical software (version 16). Results The
sensitivity of the assay with 95% probability, determined by probit
analysis, was 15 IU/µL. The assay showed a linear range of
101 IU/µL to 104 IU/µL
(R2 = 0.989). The coefficient of variation for
intra and inter assay precision of the assay, based on threshold cycle
value, ranged from 0.24 to 0.4, and from 1.94 to 3.19, respectively. The
analytical and clinical specificity was 100%. No bias in relation to
concentration between the results of 29 HCV RNA positive clinical
specimens, simultaneously tested by artus HCV LC RT-PCR reagents and
in-house reagents, was observed in the method comparison. Conclusions
The in-house 1 step TaqMan Real Time RT-PCR assay showed acceptable
performance characteristics. Our study presents the robustness and
cost-effectiveness of the method for detection and quantification of HCV
RNA.
Journal title :
Astroparticle Physics