Author/Authors :
Khodadadian, Mehdi Medical Biology Research Center - Kermanshah University of Medical Sciences, Kermanshah, Iran , Jalili, Ronak Medical Biology Research Center - Kermanshah University of Medical Sciences, Kermanshah, Iran , Bahrami, Mohammad Taher Medical Biology Research Center - Kermanshah University of Medical Sciences, Kermanshah, Iran , Bahrami, Gholamreza School of Pharmacy - Kermanshah University of Medical Sciences, Kermanshah, Iran
Abstract :
An electroanalytical method has been introduced for highly sensitive determination of
hydralazine hydrochloride (Hy-HCl) based on its oxidation at a glassy carbon electrode modified
with multiwalled carbon nanotubes (MWCNT/GCE). Studies showed that the electrochemical
oxidation of Hy-HCl was accompanied by adsorption and highly sensitive responses could be
achieved by adsorptive stripping voltammetry. The electrooxidation of Hy-HCl at MWCNT/
GCE occurred at ~32 mV which was lower than that observed at bare GCE (~52 mV). The
optimum working conditions for determination of the drug using differential-pulse adsorptive
stripping voltammetry (DPAdSV) were established. The method exhibited linear responses to
Hy-HCl in the concentration range 10-220 nM with a detection limit of 2.7 nM. The proposed
method was successfully applied to the determination of this compound in pharmaceutical
dosage forms.
Keywords :
Stripping voltammetry , Adsorption , Modified electrode , Hydralazine , Carbon nanotube