Title of article :
New approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras
Author/Authors :
Akbarzadeh, Afshin Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran , Saba, Valiallah Radiation Research Center, Faculty of Paramedicine, AJA University of Medical Sciences, Tehran, Iran , Ay, Mohammad Reza Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
Abstract :
Introduction: SURGEOSIGHT is a dedicated intra-operative hand-held gamma camera designed and manufactured by our team to
be used for lymphoscintigraphy. Although conventional gamma cameras are widely used, they lack the capability of the sentinel
lymph node (SLN) imaging in the operation room. Like conventional gamma cameras it is necessary to calibrate SURGEOSIGHT
for linearity distortion, energy and flood correction. Our main objective in this study is to develop and implement three techniques
for linearity, energy and uniformity corrections.
Methods: The camera is made up of a pixelated cesium iodide (sodium-activated) (CsI(Na)) scintillation crystal with 1 × 1 mm2
element size along with a Hamamatsu H8500 flat-panel multi-anode (64 channels) photomultiplier tube. All triple corrections
required lookup tables (LUTs). To generate all three LUTs, one uncalibrated acquisition is necessary. Linearity distortion
correction was based on localization of physical pixels of crystal and thin plate spline interpolation to estimate the amount of
distortion. Energy correction LUT was simply a photo-peak map of incoming events which was estimated using self-organizing
map technique.
Results: The results showed improvements qualitatively and quantitatively. The extrinsic energy resolution was enhanced from
50% to about 20.6%. Integral uniformity and the differential uniformity, after uniformity correction, in useful field-of-view
(UFOV) was measured 9.5% and 4.5%, respectively.
Conclusion: Results substantiated that the correction techniques guarantee uniform and accurate output of the SURGEOSIGHT
which is desirable for intra-operative localization of the sentinel lymph nodes in breast cancer and also for imaging of other
superficial tumors.
Keywords :
Uniformity correction , Energy correction , Linearity distortion correction , Self-organizing map , Intra-operative gamma camera
Journal title :
Astroparticle Physics