• Title of article

    Atomic Absorption Spectrometric Determination of Lead after Extraction and Preconcentration with 5-Br-PADAP

  • Author/Authors

    Taher, Mohammad Ali Department of Chemistry - Shahid Bahonar University - Kerman

  • Pages
    8
  • From page
    1
  • To page
    8
  • Abstract
    Lead is quantitatively retained on 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP)-ammonium tetraphenylborate with microcrystalline naphthalene or by a column method in the pH range 4.0-6.0 from a large volume of aqueous solutions of various samples with preconcentration factor of 60. After filtration, the solid mass consisting of the lead complex and naphthalene was dissolved with 5 mL of dimethylformamide and the metal was determined by atomic absorption spectrometry. The calibration curve is linear in the range of 1.9-180 mg/mL in DMF with a relative standard deviation 1.7&. The interference of a large number of anions and cations has been studied.
  • Keywords
    Alloys and biological samples , Atomic absorption spectrometry , 2-(5-Bromo-2-pyridyazo)-5-diethylamminophenol , Trace lead determination
  • Journal title
    Astroparticle Physics
  • Serial Year
    2001
  • Record number