Title of article :
The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections
Author/Authors :
Gharekhani, Vahede Department of Prosthodontics - Faculty of Dentistry - Tabriz University of Medical Sciences, Tabriz , Saber, Fariba Saleh Department of Prosthodontics - Faculty of Dentistry - Tabriz University of Medical Sciences, Tabriz , Jannatii Ataei, Soheil Department of Prosthodontics - Faculty of Dentistry - Tabriz University of Medical Sciences, Tabriz , Taghizade Motlagh, Mahsa Department of Prosthodontics - Faculty of Dentistry - Tabriz University of Medical Sciences, Tabriz , Abolfazli, Nader Department of Periodontics - Faculty of Dentistry - Tabriz University of Medical Sciences, Tabriz
Abstract :
Background. Since the misfit of crown has an important role in clinical performance of implant-supported prostheses, and
due to the impact of the settling effect on misfit, the aim of this study was to investigate the impact of torque forces on the
total lengths of narrow and short implant abutments in different internal implant‒abutment connections.
Methods. In four different implant‒abutment connections, 8 analog implants with a normal diameter (4 mm) and narrow
abutment (4.5 mm) were selected from groups of internal hex, internal octagon, morse hex 6° and morse hex 11°. Each of
them was mounted within plaster type IV, and 32 samples were obtained. Then, the amount of vertical displacement was
measured by closing the impression copings and applying torques of 20 25 and 30 Ncm. This stage was repeated for the
abutment. In the next stage, the resin pattern was built and measurements were performed after applying the torques mentioned.
Finally, after making the frame, this stage was repeated, and the settling effect was statistically analyzed with
ANOVA.
Results. In the stages of impression coping, resin pattern and final prosthesis, HEXAGONE had significantly the highest
and OCTAGONE had the lowest rates of settling, and the settling of morse hex 11° and 6° was between them.
Conclusion. Octagon implant had significantly the lowest settling in various clinical and laboratory stages by applying
different torques.
Keywords :
tightening torque , settling effect , morse hex taper , Axial displacement