• Title of article

    Method validation of methotrexate in human plasma by LC-MS technique in patients with brain tumour

  • Author/Authors

    Danafar, Hossein Zanjan Pharmaceutical Nanotechnology Research Center - Zanjan University of Medical Sciences, Zanjan - Medicinal Chemistry Dept - Zanjan University of Medical Sciences, Zanjan - Pharmaceutics Dept - Zanjan University of Medical Sciences

  • Pages
    10
  • From page
    162
  • To page
    171
  • Abstract
    Background and aims: A selective and sensitive high performance liquid chromatography-electrospray ionization mass spectrometry method has been established for determination of methotrexate in human plasma. Methods: Methotrexate was extracted from plasma with acetonitrile. The mobile phase consisted of acetonitrile-water-formic acid 74: 25: 1(v/v). Twenty μl of sample was chromatographically analyzed using a repacked ZORBAX-XDB-ODS C18 column (2.1mm×30mm, 3.5 micron). The mode of mass spectrometry was selected-ion monitoring (SIM). Results: Standard curve was linear (r= 0.998) over the concentration range of 0.1–100.0 ng/ml and showed a suitable accuracy and precision. The limit of detection (LOD) was 0.05ng/ml. Conclusion: The method is quick, easy, very steady and precise for the partition, assignment and evaluation of methotrexate in human plasma.
  • Keywords
    Methotrexate , LC-MS , Human plasma
  • Journal title
    Astroparticle Physics
  • Serial Year
    2016
  • Record number