Title of article :
Estimation of the effective dose to the radiologists during fluoroscopy or angiography of abdominal viscera
Author/Authors :
Hajizadeh Saffar, M Department of Medical Physics - Mashhad University of Medical Sciences , Nekoee, S Department of Medical Physics - Mashhad University of Medical Sciences , Bahrayni-Toosi, M.H Department of Medical Physics - Mashhad University of Medical Sciences , Sedghi, A Department of Medical Physics - Mashhad University of Medical Sciences
Abstract :
Background:
as fluoroscopy or different angiographies are usually evaluated using TL dosimeter. This
method is a time-consuming procedure therefore, radiologists are usually worried and unaware
of their exact radiation dose received during each fluoroscopy or angiography. In this study, a
new trend for quick estimation of effective dose based on measuring air-absorbed dose of the
scattered radiations at body surface of the radiologists has been introduced.
The radiation effective dose received by the radiologists during procedures such
Materials and Methods:
fluoroscopy unit (Shimatsu, model SF5010MD), and a water phantom (spherical plastic bag 27
cm in diameter filled with water up to 22.5 cm height). Radiation field was 20 × 20 cm on the
bed and X-ray tube had 1.5 and 2 mm of Aluminum as an internal and external filtration
respectively. A calibrated survey meter model RDS-110 was used to measure the scattered
radiation horizontally and vertically around the phantom at different angles and distances, in
front and behind of an apron.
Scattered radiations of abdominal viscera were simulated by a
Results:
mA, was 451 µGy/h. This value reduced to 4.45 µGy/h by passing through lead ribbons of
serigraph and to 1.2 µSv/h behind an apron. The scattered dose rate at different angles above the
bed was constant and varied by distance from the center of the radiation field. The effective
dose received by the radiologist is estimated to be about 174µGy/h, while wearing an apron and
staying 50 cm away from the patient during fluoroscopy or angiography of abdominal viscera.
The scattered dose rate at 1 m from the phantom, during fluoroscopy at 83 kVp and 1.7
Conclusion:
different angiographies of abdominal viscera, by determining scatter radiation dose at their body
surface and applying factor 0.87 for shielding effect of the body. Equivalent organ dose can also
be calculated from dose rate in air after applying factor 0.87 for deep organs or 1.1 (the mass
energy absorption coefficient ratio of water/tissue to air) for organs near the surface.
Keywords :
angiography , fluoroscopy , radiologist , Effective dose , phantom