Title of article :
Beam hardening artifacts by dental implants: Comparison of cone‑beam and 64‑slice computed tomography scanners
Author/Authors :
Esmaeili, Farzad Department of Oral Radiology - Tabriz University of Medical Sciences, Tabriz , Johari, Masume Department of Oral Radiology - Tabriz University of Medical Sciences, Tabriz , Pezhman Haddadi, Department of Oral Radiology - Tabriz University of Medical Sciences, Tabriz
Abstract :
Background: Cone beam computed tomography (CBCT) is an alternative to a computed
tomography (CT) scan, which is appropriate for a wide range of craniomaxillofacial indications. The
long‑term use of metallic materials in dentistry means that artifacts caused by metallic restorations
in the oral cavity should be taken into account when utilizing CBCT and CT scanners. The aim of
this study was to quantitatively compare the beam hardening artifacts produced by dental implants
between CBCT and a 64‑Slice CT scanner.
Materials and Methods: In this descriptive study, an implant drilling model similar to the human
mandible was used in the present study. The implants (Dentis) were placed in the canine, premolar
and molar areas. Three series of scans were provided from the implant areas using Somatom
Sensation 64‑slice and NewTom VGi (CBCT) CT scanners. Identical images were evaluated by
three radiologists. The artifacts in each image were determined based on pre‑determined criteria.
Kruskal‑Wallis test was used to compare mean values; Mann‑Whitney U test was used for
two‑by‑two comparisons when there was a statistical significance (P < 0.05).
Results: The images of the two scanners had similar resolutions in axial sections (P = 0.299). In
coronal sections, there were significant differences in the resolutions of the images produced by
the two scanners (P < 0.001), with a higher resolution in the images produced by NewTom VGi
scanner. On the whole, there were significant differences between the resolutions of the images
produced by the two CT scanners (P < 0.001), with higher resolution in the images produced by
NewTom VGi scanner in comparison to those of Somatom Sensation.
Conclusion: Given the high quality of the images produced by NewTom VGi and the lower costs
in comparison to CT, the use of the images of this scanner in dental procedures is recommended,
especially in patients with extensive restorations, multiple prostheses and previous implants.
Keywords :
Beam hardening artifacts , cone beam computed tomography , dental implants
Journal title :
Astroparticle Physics