Author/Authors :
Esmaili, G. Department of Medical Physics - Iran University of Medical Sciences, Tehran , Mahdavi, S.R. Radiation Biology Research Center & Medical Physics Department - Faculty of Medicine - Iran University of Medical Sciences, Tehran , Nikoofar, A.R. Radiation Oncology Department - Faulty of Medicine - Iran University of Medical Sciences, Tehran , Fadavi, P. Radiation Oncology Department - Faulty of Medicine - Iran University of Medical Sciences, Tehran , Ameri, A. Radiation Oncology Department - Faulty of Medicine - Shahid Beheshti University of Medical Sciences, Tehran , Nazari, V. Department of Medical Physics - Faculty of Medicine - Hamedan University of Medical Sciences, Hamedan
Abstract :
Background: Commissioning tests are recommended before implementing Intensity Modulated Radiation Therapy. This study evaluated five different plans of IMRT mocks for different modulation indices. Materials and Methods: IMRT tests were done in slab phantom for both 6MV and 15 MV photon energies using a 2D array ion chamber dosimeter. Results: The acceptance criteria were based on 3% /3mm. All tests passed the gamma criteria except for the prostate test with 15 MV photon beams and C shape test for both 6 MV and 15 MV energies. The Ratio passing points for Multi target test in 6 MV and 15 MV plans were 93.1% and 91.5%, respectively. The results of the prostate tests with 3%/3mm was 92.3% for 6MV and 91.2% for 15 MV in 4%/4 mm tolerance limit. For Head and Neck test with the same gamma criteria, the percentages of the points were 93.2 and 94 for 6 MV and 15 MV plans, respectively. For C shape tests, the used gamma criteria were 4%/ 4 mm. The ratio passing points were 94.9% and 94.3% for 6MV and 15MV plans, respectively. However, C shape hard test could not pass the gamma criteria of 4%/ 4mm for 6MV. Conclusion: results showed that by increasing the complexity of the IMRT plan, the verification test must be done in a more strict-manner, because a small change in dose delivery can cause a large discrepancy between planned and real dosimetry and may produce hot spots in organs or a cold spot in the target volume.
Keywords :
Pre- treatment IMRT , 2D array , IMRT mock , Gamma evaluation