Title of article :
Investigation of LINAC Structural Effects on Photoneutron Specified Parameters Using FLUKA code
Author/Authors :
Ashrafinia, Mohammad Department of Medical Radiation Engineering - Science and Research Branch Islamic Azad University, Tehran, Iran , Hadadi, Ashghar Department of Medical Radiation Engineering - Science and Research Branch Islamic Azad University, Tehran, Iran , Sardari, Dariush Department of Medical Radiation Engineering - Science and Research Branch Islamic Azad University, Tehran, Iran , Saeedzadeh, Elham Department of Medical Radiation Engineering - Science and Research Branch Islamic Azad University, Tehran, Iran
Abstract :
Introduction: The utilization of high-energy photons in the medical linear accelerator can lead to
photoneutron production. This study aimed to evaluate the effect of the physical components of the head,
including flattening filter (FF) and multileaf collimator (MLC), as well as the dependence of therapeutic field
size on the photoneutron spectrum, dose, and flux.
Material and Methods: The present study reported the simulation of the fundamental linac head components
of the Varian Clinac 2100 performing in X-ray mode with 18 MV energy by the FLUKA code. The
percentage depth dose and lateral dose profile were measured using a PTW thimble chamber to ensure the
simulation reliability.
Results: Photoneutron spectrum analysis indicated that neutrons with highest relative biological effectiveness
were delivered to the phantom surface, and opening the field from 0×0 to 40×40𝑐𝑚2 shifted the spectrum by
24.545% to the higher energies. The target and the vicinity parts played the most prominent roles in neutron
contamination. The relationship between the field size and the photoneutron dose was non-linear, and it
reached a peak of 20×20 cm2. Although using small fields formed by the MLC contribute to a lower dose
compared to those shaped by the jaws, MLC-equipped machines result in 21.98% higher dose. Moreover, the
flattening filter removal unexpectedly increased the isocenter photoneutron dose by 11.63%. This undesirable
dose can be up to 2.54 mSv/Gy for the reference field at the isocenter while the out-of-field dose is about 0.5
mSv/Gy for most of the field dimensions.
Conclusion: As a result, it is critical to consider this unwanted absorbed dose, which is seriously influenced
by the implemented therapeutic conditions.
Keywords :
Fast Neutrons , Linac , Monte Carlo , FLUKA
Journal title :
Astroparticle Physics