Title of article :
Pencil‑Beam Fluence Evaluation Based on Monte Carlo Simulations Algorithm of High Energetic Treatment Photons
Author/Authors :
donya, hossam Department of Physics - Faculty of Science - King Abdulaziz University, Jeddah 21589, Saudi Arabia
Pages :
6
From page :
81
To page :
86
Abstract :
Recently, the models used for this purpose have been the point kernel, pencil‑beam, and collapsed cone superposition/convolution combination models. Methods: In this study, it is aimed to determine point/pencil‑beam kernels to be used in dose calculations. For this purpose, tallying pencil‑beam fluence based on Monte Carlo (MC) simulations is investigated by scoring a volume of interest centered in a cubic water phantom at a depth of 10 cm and the irradiated field of 10 cm × 10 cm. The fluence is calculated for each mono‑energetic photon ranging from 0.25–6 MV at increments of 250 keV. Results: The output of the four fluence kernels along the depth Z around the central axis is categorized for both the primary and secondary photons and electrons. Here, a database for pencil‑beam kernels is established for each category. For validation purposes, other MC simulations are carried out for fluence calculations as produced by the predetermined poly‑energetic spectra for a Varian 6 MV linear accelerator. Conclusions: The net results show a good fit of the two convoluted fluence spectra quantities for both mono‑energetic and poly‑energetics‑based simulations except little singular peaks.
Keywords :
Electron fluence , Monte Carlo simulation , pencil‑beam kernel , photon fluence , solid‑state dosimeters
Serial Year :
2018
Record number :
2492363
Link To Document :
بازگشت