Title of article :
Causes of Fracture at Catheter of Totally Implantable Venous Access Port: A Systematic Review
Author/Authors :
Kridis, Wala Ben Department of Medical Oncology - Habib Bourguiba University Hospital - Tunisia , Toumi, Nabil Department of Medical Oncology - Habib Bourguiba University Hospital - Tunisia , Khanfir, Afef Department of Medical Oncology - Habib Bourguiba University Hospital - Tunisia
Pages :
4
From page :
686
To page :
689
Abstract :
A totally implantable venous access port (TIVAP) plays a crucial role in the treatment of patients in oncology. Catheter fracture is a serious complication with an estimated incidence of 0, 1% - 1%. The objective of this systematic review is to analyze the mechanism of TIVAP fracture to make physicians aware of this fatal entity. A search of the literature between 1980 and 2019 was conducted using PubMed, Ovid, MEDLINE, and Cochrane Systematic Review databases. The search identified 18 case reports and 8 retrospective studies. Fracture of the middle part of the catheter may be induced by constant compression of the catheter between the first-rib and clavicle, which is called the pinch-off syndrome. Catheter fracture at the port-catheter junction may be caused by extrinsic compression near the port-catheter junction combined with material fatigue due to repeated bending of the catheter with shoulder movement. There is no specific cause for the fracture of a catheter tip. An annual chest X-ray is recommended for the early detection of TIVAP catheter fracture. Percutaneous endovascular retrieval of a dislodged Port-A catheter is both safe and effective.
Keywords :
Totally implantable venous access port , Fracture , Mechanism , Oncology , Pinch off syndrome
Journal title :
Acta Medica Iranica
Serial Year :
2019
Record number :
2497610
Link To Document :
بازگشت