Rohani, S. A Department of Medical Physics - Tehran University of Medical Sciences, Tehran, Iran , Mahdavi, S. R Radiation biology research center & medical Physics department - faculty of medicine - Iran University of Medical Sciences, Tehran, Iran , Mostaar, A Department of Medical Physics - School of Medicine - Shahid Beheshti University of Medical Sciences, Tehran, Iran , Ueltzhöffer, S Department of Clinic for Radiotherapy and RadioOncology - Medical Faculty Mannheim of the University of Heidelberg, Heidelberg, Germany , Mohammadi, R Department of Medical Physics - Iran University of Medical Sciences, Tehran, Iran , Geraily, Gh Department of Medical Physics - Tehran University of Medical Sciences, Tehran, Iran
Abstract :
Background: Before treatment planning and dose delivery, quality assurance of
multi-leaf collimator (MLC) has an important role in intensity-modulated radiation
therapy (IMRT) due to the creation of multiple segments from optimization process.
Objective: The purpose of this study is to assess the quality control of MLC
leaves using EBT3 Gafchromic films.
Material and Methods: Leaf Position accuracy and leaf gap reproducibility
were checked with Garden fence test. The garden fence test consists of 5 thin bands
A) 0.2 Cm width spaced at 2 Cm intervals and B) 1 Cm width spaced at 1 Cm intervals.
Each leaf accuracy was analyzed with measuring the full-width half-maximum
(FWHM). Maximum and average leaf transmission were measured with gafchromic
EBT3 films from Ashland for both 6 MV and 18 MV beams.
Results: Leaf positions were found to be in a range between 1.78 – 2.53 mm,
instead of nominal 2 mm for the test A and between 9.09 – 10.36 mm, instead of
nominal 10 mm for the test B. The Average radiation transmission of the MLC was
noted 1.79% and 1.98% of the open 10x10 Cm2 field at isocenter for 6 MV and 18
MV beams, respectively. Maximum radiation transmission was noted 4.1% and 4.4%
for 6 MV and 18 MV beams, respectively.
Conclusion: In this study, application of gafchromic EBT3 films for the quality
assurance of Euromechanics multileaf collimator was studied. Our results showed
that the average leaf leakage and positional accuracy of this type of MLC were in the
acceptance level based on the Protocols.
Keywords :
Leaf End Transmission , Leaf Transmission , Garden Fence Test , Mechanical Test , Multileaf Collimator