Zaman, M. U PET/CT Imaging Services - Department of Radiology - Aga Khan University Hospital, Karachi, Pakistan , Fatima, N PET/CT Imaging Services - Department of Radiology - Aga Khan University Hospital, Karachi, Pakistan , Mehdi, S. U PET/CT Imaging Services - Department of Radiology - Aga Khan University Hospital, Karachi, Pakistan , Sami, M Allmed Solution, Karachi, Pakistan , Khan, Kh PET/CT Imaging Services - Department of Radiology - Aga Khan University Hospital, Karachi, Pakistan , Khan, G PET/CT Imaging Services - Department of Radiology - Aga Khan University Hospital, Karachi, Pakistan
Abstract :
Background: Compact cyclotrons are getting popular to fulfill enormous current
demands of PET tracers. Aga Khan University Hospital, Karachi, Pakistan has
acquired the first smallest footprint of BG-75 Generator for 18FDG-based PET/CT
clinical imaging. We are sharing our experience of BG-75 in the first year (December
2015-November 2016) after commissioning.
Material and Methods: BG-75 Generator (ABT, USA) was installed in available
space without major design modification. It has a self-shielded mini-cyclotron
(7.5 MeV proton beam, positive ion with current < 5μA) to produce [18F]F-, an
automated card-based radiochemistry module to produce 18FDG and automated QC
module to perform tests upon each batch of 18FDG. Data were collected for yields of
[18F]F-, 18FDG, QC and radiation safety parameters.
Results: Total 545 runs in 167 days (3±01 runs/day) were made. Average yield
with 60 minute bombardment using 4.5 μA current was 37 mCi and 21 mCi for [18F]
F- and 18FDG, respectively. Total 29 runs had chemistry or QC failures and were
discarded. Remaining 516 batches were used to perform imaging upon 1370 patients
(8±03 patients/day). Radiation dose in BG-75 suite and effective doses to 02 operators
were within statutory limits.
Conclusion: BG-75 Generator provides safe, dependable and sustainable supply
of 18F-Fluoride for 18FDG and other low-volume clinical PET imaging. Its compactness
and automation need minimal space and manpower. Radiation dose rate in
cyclotron suite and personal dosimetry were also found within safe limits. Its Doseon-
Demand workflow offers a new concept of Personalized Dose Preparation which
is currently not possible with a conventional cyclotron.