Enhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode
Author/Authors :
Zeghari, A Laboratory of Nuclear Physics - Faculty of Sciences - Mohammed V University, Rabat, Morocco , Saaidi, R Laboratory of Nuclear Physics - Faculty of Sciences - Mohammed V University, Rabat, Morocco , Cherkaoui El Moursli, R Laboratory of Nuclear Physics - Faculty of Sciences - Mohammed V University, Rabat, Morocco
Abstract :
Purpose: In the last years, some studies investigated dosimetric benefits of a
free flattening filter for the photon mode in the radiotherapy field. This study aims to
provide a theoretical study published and analysis of basic dosimetric properties for a
Saturne 43 Linac head to implement free flattening filter beams clinically.
Material and Methods: This is the first Monte Carlo study for the head of
Saturne 43 with replacement flattening filter mode investigating beam dosimetric
characteristics, including central axis absorbed doses, beam profiles and photon
energy spectra. The later ones were analyzed for flattening filter and replacement flattening
filter beams using BEAMnrc and DOSXYZnrc Monte Carlo codes for 10 × 10
cm2, 5 × 5 cm2 and 2 × 2 cm2 square field sizes.
Results: A 3.94-fold increase of dose rate and electron contaminating increased
by 246.4 % with the replacement flattening filter mode for field size of 10 × 10 cm2.
Reduction was made by replacement flattening filter beam in the peripheral dose up
to 30%, and the time was reduced more than 50 %.
Conclusion: Results obtained from our study revealed that some characteristic
dosimetries such as the maximum increase in depth dose rate, decrease in out-ofdepth
dose, and reducing time can be beneficial for the unflattened beam to be used
in the radiotherapy for the advanced techniques.
Keywords :
Stereotactic Radiation , Radiation Dosimetry , Radiotherapy , Monte Carlo Method
Journal title :