Title of article :
Validation of the Gel & Wax Boluses and Comparison of their Dosimetric Performance with Virtual Bolus
Author/Authors :
Verma, T. R Department of Radiotherapy - King George’s Medical University, UP, Lucknow, India , Painuly, N. K Department of Radiotherapy - King George’s Medical University, UP, Lucknow, India , Tyagi, M Department of Radiology & Radiotherapy, UCMS, Delhi, India , Johny, D Department of Radiotherapy - King George’s Medical University, UP, Lucknow, India , Gupta, R Department of Radiotherapy - King George’s Medical University, UP, Lucknow, India , Bhatt, M. L. B Department of Radiotherapy - King George’s Medical University, UP, Lucknow, India
Abstract :
Background: In general, radiotherapy treatment planning is performed using the
virtual bolus. It is necessary to investigate physical bolus in comparison to virtual
one.
Objectives: In the present study, first, radiological properties of superflab Gel
bolus and Paraffin wax bolus was investigated in terms of their relative electron density.
Then, dosimetric performance of both the bolus (i.e. Gel and Parafin wax) was
compared with Virtual bolus.
Material and Methods: In This experimental study, the radiological property
of Wax and Gel boluses was investigated using two methods. In one, the relative
electron density of both the Gel and Wax boluses was calculated by measuring their
linear attenuation coefficient where in another method relative electron density was
calculated by recording their CT No directly from their CT scan. Later CT scan of
solid water slab phantom (dimension 30x30x15 cm3), with physical boluses (i.e. Gel
and Wax bolus) of appropriate thicknesses required to deliver a dose of 200 cGy at
Dmax using 4 MV, 6 MV and 15 MV photon beams, was taken. These CT data sets were
retrieved to TPS. A plan was done to deliver a dose of 200 cGy at Dmax using Single
4 MV, 6 MV and 15 MV photon beams. Dose at depths Dmax, 1 cm, 2 cm, 3 cm, 4 cm
and 5 cm was recorded. Using this similar method, doses at depths viz Dmax, 1 cm, 2
cm, 3 cm, 4 cm and 5 cm was recorded for the Gel and Wax boluses. The differences
in dose of gel and wax bolus from virtual bolus were recorded for comparison of their
dosimetric performance.
Results: The measured (calculated) relative electron density of wax and Gel bolus
was found to be 0.958 (0.926) and 0.923 (0.907), respectively. Variation in dosimetric
performance of Gel and Wax with reference to Virtual bolus was studied. However, on
average, Gel bolus was more consistent with virtual bolus.
Conclusion: To avoid any dose difference between, delivered (using physical bolus)
and planned (using virtual bolus), the physical boluses should be investigated for
their dosimetric performance in comparison to virtual bolus. The results obtained and
methodology used in this study can be applied in routine radiotherapy practices.
Keywords :
Waxes , Imaging , X-Ray Computed , Tomography , Paraffin
Journal title :
Journal of Biomedical Physics and Engineering