Comparison of the 2-D Dose Distribution Calculated by Planning System and Measured by Gafchromic Film Physical Dosimetry for 60Co and 192Ir Brachytherapy Sources
Author/Authors :
Gholami ، M H Department of Medical Radiation Engineering - Islamic Azad University, Science and Research Branch , Sadeghi ، M Medical Physics Department - Finetech in Medicine Research Center, School of Medicine - Iran University of Medical Sciences , Babapour Mofrad ، F Department of Medical Radiation Engineering - Islamic Azad University, Science and Research Branch , Mohammadi ، M Department of Medical Physics - Royal Adelaide Hospital, School of Physical Sciences - University of Adelaide
Abstract :
Background: High Dose Rate (HDR) brachytherapy sources with high photon energy have been widely used in treating tumors. Dosimetric parameter of these brachytherapy sources should be determined according to the AAPM TG43 recommendation. Gafchoromic films are reliable tools for this evaluation. Objective: The aim of this study is to evaluate and compare dose accuracy of the twobrachytherapy sources in a dedicated phantom.Material and Methods: In this analytical study, two common sources, including Cobalt and Iridium, were loaded into the dedicated phantom. The twodimensional dose distribution around the source was calculated by TPS system for certain activities and geometries around the sources. Then, the experimental dose measured by Gafchromic film dosimetry was reported for different angles ranging from 0 to 180 degrees. Results: The difference between calculated and measured doses was less than 6 percent (5 to +6 percent) for all of the channels and angles. These errors are smaller and mainly more than zero (Dfilm gt;DTPS) for angles less than 20 and larger than 110 degrees. There is no statistically significant discrepancy in dose calculation by treatment planning system. Conclusion: Although the estimated error in dose calculation is not significant, there is still an opportunity to increase the treatment precision. The correlation between the error and the angle should be considered in further plans of brachytherapy. The present study showed comparable errors compared to results of other research studies.
Keywords :
Brachytherapy , Film dosimetry , Cobalt , 60 , Iridium , 192
Journal title :
Journal of Biomedical Physics and Engineering
Journal title :
Journal of Biomedical Physics and Engineering