Title of article :
Impact of reconstruction method on quantitative parameters of 99mTc-TRODAT-1 SPECT
Author/Authors :
Ghaedian, Tahereh Nuclear Medicine and Molecular Imaging Research Center - Namazi Teaching Hospital - Shiraz University of Medical Sciences, Iran , Saleki, Leila Nuclear Medicine and Molecular Imaging Research Center - Namazi Teaching Hospital - Shiraz University of Medical Sciences, Iran , Ebadi, Fatemeh Nuclear Medicine and Molecular Imaging Research Center - Namazi Teaching Hospital - Shiraz University of Medical Sciences, Iran , Rakhsha, Abbas Department of Neurosurgery - Shiraz University of Medical Sciences, Iran
Abstract :
Introduction: Quantitative evaluation is recommended to improve diagnostic ability and serial assessment of dopamine
transporter (DAT) density scans. We decided to compare the ordered subsets expectation-maximization (OSEM) with filtered
back-projection (FBP), and to investigate the impact of different iteration and cut-off frequencies on SBR values.
Methods: We retrospectively examined 27 consecutive patients. SPECT reconstruction was performed using OSEM and FBP
with Chang’s attenuation correction (AC). Iterative reconstruction parameters were used with different iterations ranging from
2, 4, 6, 8, and 10 with fixed 10 subsets and different subsets including 5, 10 and 15 with fixed 6 iterations. Reconstruction with
FBP were performed with different critical cut-off frequencies of 0.3, 0.4 and 0.5.
Results: Comparing SBR derived by OSEM reconstruction with 10 subsets but different iterations revealed statistically
significant intraclass correlation (ICC) in both right and left side. There is also no significant difference between different
OSEM reconstruction with different subsets and ICC was excellent in all patients. ICC for FBP reconstruction with different
cut-off frequency revealed good ICC in all patients. However, lower degree of SBR showed higher decrease in ICC with
insignificant and poor correlation in patients with SBR<0.2. While comparing OSEM and FBP, good correlation was observed
in total patients, there was poor correlation between these reconstruction methods in lower SBR values.
Conclusion: Our study showed that change in FBP reconstruction parameters can greatly impact the SBR value of 99mTcTRODAT-1, especially in patients with more severe disease. However, OSEM reconstruction revealed better reproducibility
for SBR using different iterations.
Keywords :
SPECT reconstruction , 99mTc-TRODAT-1 , Ordered subsets expectation-maximization , Filtered back-projection
Journal title :
Iranian Journal of Nuclear Medicine