Title of article :
Polarized Mueller Matrix Analytical Model for Glucose Measurement in Vitro
Author/Authors :
FIRDOUS, Shamaraz Pakistan Institute of Engineering and Applied Sciences (PIEAS - Department of Physics and Applied Mathematics , IKRAM, Masroor Pakistan Institute of Engineering and Applied Sciences (PIEAS - Department of Physics and Applied Mathematics
From page :
149
To page :
155
Abstract :
In this study we present a noninvasive and nonradioactive glucose in vitro measurement technique using the Muellerpolarization analytical model. The glucose solution and aqueous humor phantoms are illuminated with a He-Ne (632.8 nm) laser.This in vitro study provides a base and reference data for in vivo eye glucose monitoring. The 16 Mueller polarization matrix elementimages give information about eye absorption and scattering, and changes in polarization patterns to measure glucose concentration.The interior part of the eye (aqueous humor) contains a birefringent scattering medium for glucose measurement. The Muellermatrix polarimetric technique for simultaneous extraction of rotating linear and circular polarized light is discussed theoretically andexperimentally implemented. This method requires calculation of the Mueller matrix for different orientations of polarized lightthrough turbid samples. The results are in good agreement with those in the literature
Keywords :
Polarization , glucose sensor , birefringence , depolarization , aqueous humor , Mueller matrix polarimeter
Journal title :
Turkish Journal of Medical Sciences (TJMS)
Journal title :
Turkish Journal of Medical Sciences (TJMS)
Record number :
2528982
Link To Document :