Author/Authors :
Azizi, Arash Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Prati, Carlo Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Schiavon, Riccardo Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Michelle Fitzgibbon, Raquel Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Pirani, Chiara Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Iacono, Francesco Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Andrea Pelliccioni, Gian Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Spinelli, Andrea Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Zamparini, Fausto Department of Biomedical and Neuromotor Sciences DIBINEM - Endodontic Clinical Section - Faculty of Dentistry - University of Bologna, Bologna, Italy , Puddu, Pietro Department of Engineering Enzo Ferrari (DIEF) - University of Modena and Reggio Emilia, Modena, Italy , Bolelli, Giovanni Department of Engineering Enzo Ferrari (DIEF) - University of Modena and Reggio Emilia, Modena, Italy , Generali, Luigi Department of Surgery - Medicine, Dentistry and Morphological Sciences with Transplant Surgery, Oncology and Regenerative Medicine Relevance (CHIMOMO) - University of Modena and Reggio Emilia, Modena, Italy
Abstract :
To define surface, mechanical, microstructural and metallurgical features of conventional OneShape (OShape) and heat-treated OneCurve (OCurve) nickel-titanium instruments.
Methods: Instruments were analysed by scanning electron microscopy (SEM) on new instruments and after
simulated clinical use (SCU). Cyclic fatigue testing was performed and the number of cycles to fracture (NCF)
and the length of the fractured instruments were measured (Mann-Whitney test). Fractured instruments
during cyclic fatigue testing were then inspected by SEM fractographic analysis. Field emission gun scanning
scanning electron microscopy (FEG-SEM), energy-dispersive X-ray spectroscopy (EDX) and micro-Raman
spectroscopy were used to assess alloy surface chemistry. Focused ion beam (FIB) was performed to analyse the oxide layer on the surface of OCurve before and after SCU. X-Ray diffraction (XRD), metallographic
evaluation and differential scanning calorimetry (DSC) were used to determine martensitic/austenitic phase
transformation temperatures.
Results: SEM observations on new instruments revealed a smooth regular surface with flattened milling
grooves. No wear features were detected after SCU. OCurve exhibited a higher cyclic fatigue resistance
(P<0.05), slower crack propagation and a surface layer of TiO2
. Metallographic analysis and XRD showed the
prevalence of martensitic grains on OCurve instruments that were stable at body temperature as confirmed
by DSC analysis. Furthermore, DSC demonstrated a shift in the temperature transformation ranges suggesting an increase of martensite phase in autoclaved OCurve instruments.
Conclusion: Heat treatment processes were confirmed as a valid enhancement of the properties of the new
generation NiTi instruments. OCurve presented a significant improvement over OShape regarding both mechanical and metallurgical characteristics.
Keywords :
Differential scanning calorimetry , martensitic instruments , OneCurve , OneShape , wear analysis , X-Rays spectrometry