Title of article :
Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife
Author/Authors :
Allahverdi Pourfallah, Tayeb mazandaran university of medical sciences - Faculty of Medicine - Biochemistry and Biophysics Dept, ايران , Allahverdi, Mahmoud tehran university of medical sciences tums - Faculty of Medicine - Medical Physics Dept, تهران, ايران , Zahmatkesh, Mohammad Hasan Novin Medical Radiation Institute, ايران
From page :
1
To page :
8
Abstract :
Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with each other.Materials and Methods The head phantom was a custom-built 16 cm diameter plexiglas sphere. Inside the phantom, there were two cubic cutouts for inserting the gel vials and inhomogeneities. Following irradiation with the Gamma Knife unit, the polymer gel dosimeters were scanned with a 1.5 T MRI scanner. For the purpose of simulation the simplified channel of 60Co source ofGamma Knife BEAMnrc and for extracting the 3D dose distribution in the phantom, DOSXYZnrc codes were used.Results Within high isodose levels ( 80%), there are dose differences higher than 7%, especially between air inserted and PTFE inserted phantoms, which were obtained using both simulation and experiment. This means that these values exceed the acceptance criterion of conformal radiotherapy and stereotactic radiosurgery (i.e ., within some isodose levels , less than 93% of prescription dose are delivered to the target).Conclusion: The discrepancies observed between the results obtained from heterogeneous and homogeneous phantoms suggest that Leksell Gamma Knife planning system (LGP) predictions which assume the target as a homogeneous material must be corrected in order to take care ofthe air- and bone-tissue inhomogeneities.
Keywords :
Dose Profile , Gamma Knife , Inhomogeneity , Monte Carlo , Polymer Gel Dosimeter
Journal title :
Iranian Journal of Medical Physics (IJMP)
Journal title :
Iranian Journal of Medical Physics (IJMP)
Record number :
2570820
Link To Document :
بازگشت