Title of article :
Validation of Datascope Accutorr PlusTM using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines
Author/Authors :
Khawaja, Rajab Ali King Saud University - King Khalid University Hospital - Department of Family and Community Medicine, Saudi Arabia , Qureshi, Riaz King Saud University - King Khalid University Hospital - Department of Family and Community Medicine, Saudi Arabia , Mansure, Abdul Hay King Saud University - King Khalid University Hospital - Department of Family and Community Medicine, Saudi Arabia , Yahya, Mohammad Emran King Saud University - King Khalid University Hospital - Department of Family and Community Medicine, Saudi Arabia
From page :
1
To page :
5
Abstract :
To assure the accuracy and reliability of blood pressure measurement by noninvasive blood pressure monitor using Datascope Accutorr Plus (Paramus, NJ, USA) against mercury manometer, among adult male participants. Method: Eighty participants from a family physician’s office at a teaching hospital were recruited. One hundred and sixty measurements of blood pressure were performed according to BHS technique protocol. Statistical analysis: Descriptive analysis was done according to the AAMI and BHS protocol guidelines. The limits of agreement between the device and the standard were plotted using the method of Bland and Altman plot. Results: The mean difference ± SD between the Datascope Acutorr Plus and observer was 2.7 ± 5.2 mm Hg and 1.5± 3.26 mm Hg for systolic and diastolic blood pressure, respectively. Datascope Acutorr Plus obtained A/A grading for both systolic and diastolic blood pressure.
Keywords :
AAMI , BHS , BP devices , Datascope Acutorr plus , Validation
Journal title :
journal of the saudi heart association
Journal title :
journal of the saudi heart association
Record number :
2574242
Link To Document :
بازگشت