Title of article :
A Monte Carlo Study of the Photon Spectrum due to the Different Materials Used in the Construction of Flattening Filters of LINAC
Author/Authors :
Estepa Jiménez, J. S Universidad Nacional de Costa Rica (UNA) - Apartado Postal - Heredia, Costa Rica , Díaz Lagos, M Universidad Pedagogica y Tecnologica de Colombia - Boyaca, Colombia , Martinez-Ovalle, S. A Universidad Pedagogica y Tecnologica de Colombia - Boyaca, Colombia
Abstract :
Different types the spectrum of photons were studied; they were emitted from the flattening filter of a LINAC Varian 2100 C/D
that operates at 15 MV. The simplified geometry of the LINAC head was calculated using the MCNPX code based on the studies of
the materials of the flattening filter, namely, SST, W, Pb, Fe, Ta, Al, and Cu. These materials were replaced in the flattening filter to
calculate the photon spectra at the output of this device to obtain the spectrum that makes an impact with the patient. The different
spectra obtained were analyzed and compared to the emission from the original spectra configuration of the LINAC, which uses
material W. In the study, different combinations of materials were considered in order to establish differences between the use
of different materials and the original material, with the objective of establishing advantages and disadvantages from a clinical
standpoint.
Keywords :
LINAC , Construction , Photon , Carlo
Journal title :
Computational and Mathematical Methods in Medicine