Title of article :
False Positive Transit Time Flowmetry Graft Failure in Multivessel Coronary Spasm following Off-Pump Coronary Artery Bypass Grafting
Author/Authors :
Kassimis, George Cardiology Department - Gloucestershire Hospitals NHS Foundation Trust, Cheltenham, UK , Krasopoulos, George Oxford Heart Centre - Oxford University Hospital, Oxford, UK
Abstract :
Intraoperative Transit Time Flowmetry is currently recommended to assess graft patency during coronary artery bypass grafting
(CABG), especially in presence of haemodynamic instability or inability to wean the patient from cardiopulmonary bypass, new
regional wall motion abnormalities, or significant ventricular arrhythmias. The VeriQ system is one of the currently available
systems, which detects imperfections that may be corrected by graft revision. In this case report, multivessel coronary spasm (CS)
post-CABG interferes with these intraoperative parameters misleading initially into false results. Cardiac surgeons should bear in
mind the limit of VeriQ in distinguishing between graft failure and CS. Angiography may be considered in patients with decreased
graft flow despite revision of anastomosis and vasodilatory treatment for the definitive diagnosis.
Keywords :
Spasm , Off-Pump Coronary , Artery Bypass Grafting
Journal title :
Case Reports in Cardiology