Title of article :
A New Measure of Pulse Rate Variability and Detection of Atrial Fibrillation Based on Improved Time Synchronous Averaging
Author/Authors :
Ding, Xiaodong Shanghai University of Traditional Chinese Medicine - Shanghai, China , Wang, Yiqin Shanghai University of Traditional Chinese Medicine - Shanghai, China , Hao, Yiming Shanghai University of Traditional Chinese Medicine - Shanghai, China , Lv, Yi Shanghai University of Traditional Chinese Medicine - Shanghai, China , Chen, Rui Shanghai University of Traditional Chinese Medicine - Shanghai, China , Yan, Haixia Shanghai University of Traditional Chinese Medicine - Shanghai, China
Abstract :
Pulse rate variability monitoring and atrial fibrillation detection algorithms have been widely used in wearable devices,
but the accuracies of these algorithms are restricted by the signal quality of pulse wave. Time synchronous averaging is a powerful
noise reduction method for periodic and approximately periodic signals. It is usually used to extract single-period pulse waveforms,
but has nothing to do with pulse rate variability monitoring and atrial fibrillation detection traditionally. If this method is improved
properly, it may provide a new way to measure pulse rate variability and to detect atrial fibrillation, which may have some potential
advantages under the condition of poor signal quality. Objective. The objective of this paper was to develop a new measure of pulse
rate variability by improving existing time synchronous averaging and to detect atrial fibrillation by the new measure of pulse rate
variability. Methods. During time synchronous averaging, two adjacent periods were regarded as the basic unit to calculate the
average signal, and the difference between waveforms of the two adjacent periods was the new measure of pulse rate variability.
3 types of distance measures (Euclidean distance, Manhattan distance, and cosine distance) were tested to measure this
difference on a simulated training set with a capacity of 1000. The distance measure, which can accurately distinguish regular
pulse rate and irregular pulse rate, was used to detect atrial fibrillation on the testing set with a capacity of 62 (11 with atrial
fibrillation, 8 with premature contraction, and 43 with sinus rhythm). The receiver operating characteristic curve was used to
evaluate the performance of the indexes. Results. The Euclidean distance between waveforms of the two adjacent periods
performs best on the training set. On the testing set, the Euclidean distance in atrial fibrillation group is significantly higher than
that of the other two groups. The area under receiver operating characteristic curve to identify atrial fibrillation was 0.998. With
the threshold of 2.1, the accuracy, sensitivity, and specificity were 98.39%, 100%, and 98.04%, respectively. This new index can
detect atrial fibrillation from pulse wave signal. Conclusion. This algorithm not only provides a new perspective to detect AF but
also accomplishes the monitoring of PRV and the extraction of single-period pulse wave through the same technical route,
which may promote the popularization and application of pulse wave.
Keywords :
Synchronous , Fibrillation , ECG
Journal title :
Computational and Mathematical Methods in Medicine