Title of article :
The effect of metal artefact on the design of custom 3D printed acetabular implants
Author/Authors :
Di Laura, Anna The Royal National Orthopaedic Hospital - Brockley Hill - Stanmore - London , UK , Henckel, Johann The Royal National Orthopaedic Hospital - Brockley Hill - Stanmore - London , UK , Hothi, Harry The Royal National Orthopaedic Hospital - Brockley Hill - Stanmore - London , UK , Hart, Alister J. The Royal National Orthopaedic Hospital - Brockley Hill - Stanmore - London , UK , Wescott, Robert Institute of Orthopaedics and Musculoskeletal Science - University College London - London, UK
Abstract :
3D Printed custom-made implants constitute a viable option in patients with acetabular Paprosky III
defects. In these patients, needing complex hip revision surgery, the appreciation of the bony defect is crucial to
assure stable fixation of the customised implant, often intended to replace a failed one. We aimed to understand
the effect of metal artefact on the design of customised implants.
Methods: 26 patients with massive acetabular defects were referred, between May 2016 and September 2018, to
our institution classified as “un-reconstructable”by other hospitals. They all received custom 3D-printed acetabular
cups. A subset of them underwent two-stage revision surgery due to infection. We then extended the two-stage
procedure to the cases where metal artefacts were significantly affecting the reading of the CT scans.
CT scans of patients’pelvises were taken pre and post-implant removal. We assessed for changes in bony shape
and volume of the pelvis using 3D imaging software and quantified the effect on implant design with CAD
software.
Results: Eight (out of 26) patients (31%) underwent two-stage revision surgery. The CT bony reconstructions
between the two timepoints changed in all cases. The changes were mostly associated to the shape and
distribution of the acetabular defects. Three of these cases (37.5%) showed a remarkable difference in the
remaining bone that led to a change in implant design. So far, there has been no difference in the clinical outcome
between the patients who underwent single (n= 18) and two-stage surgery (n= 8).
Conclusions: The shape of the acetabulum reconstructed from CT data is potentially altered by metal artefact and
bone excised during removal of the failed component. For “end-of-road”acetabular reconstruction, we recommend
surgeons consider the use of two-stage surgery to enable a reliable fitting of the complex shape of 3D-printed
implants.
Keywords :
Paprosky acetabular classification , Revision hip surgery , Metal artefact , Computed tomography , Custom 3D printed implants
Journal title :
3D Printing in Medicine