    New patient-controlled abdominal compression method in radiography: radiation dose and image quality

  • Author/Authors

    Piippo-Huotari, Oili Department of Radiology - Faculty of Health Sciences - Örebro University, Sweden , Norrman, Eva Department of Medical Physics - Faculty of Medicine and Health - Örebro University, Sweden , Anderzén-Carlsson, Agneta University Health Care Research Center - Faculty of Medicine and Health - Örebro University, Sweden , Geijer, Håkan Department of Radiology - Faculty of Medicine and Health - Örebro University, Sweden

  • Pages
    8
  • From page
    1
  • To page
    8
  • Abstract
    Background The radiation dose for patients can be reduced with many methods and one way is to use abdominal compression. In this study, the radiation dose and image quality for a new patient-controlled compression device were compared with conventional compression and compression in the prone position. Purpose To compare radiation dose and image quality of patient-controlled compression compared with conventional and prone compression in general radiography. Material and Methods An experimental design with quantitative approach. After obtaining the approval of the ethics committee, a consecutive sample of 48 patients was examined with the standard clinical urography protocol. The radiation doses were measured as dose-area product and analyzed with a paired t-test. The image quality was evaluated by visual grading analysis. Four radiologists evaluated each image individually by scoring nine criteria modified from the European quality criteria for diagnostic radiographic images. Results There was no significant difference in radiation dose or image quality between conventional and patient-controlled compression. Prone position resulted in both higher dose and inferior image quality. Conclusion Patient-controlled compression gave similar dose levels as conventional compression and lower than prone compression. Image quality was similar with both patient-controlled and conventional compression and was judged to be better than in the prone position.
  • Keywords
    Compression , image quality , radiation dose , radiography , X-ray
  • Journal title
    Acta Radiologica Open
  • Serial Year
    2018
  • Record number

    2620047