Mussmann, Bo Department of Radiology - Odense University Hospital, Odense, Denmark , Andersen, Poul Erik Department of Radiology - Odense University Hospital, Odense, Denmark , Torfing, Trine Department of Radiology - Odense University Hospital, Odense, Denmark , Overgaard, Søren Department of Clinical Research - University of Southern Denmark, Odense, Denmark
Abstract :
Background
Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current CT workstations
Purpose
To test the between-scan agreement and reliability of custom segmentation software for BMD measurements adjacent to cemented and uncemented acetabular cups in dual-energy CT (DECT).
Material and Methods
Twenty-four male patients with total hip arthroplasty were scanned and rescanned using 130-keV virtual monochromatic DECT images. Hemispherical regions of interest were defined slice-by-slice and BMD was calculated around the acetabular cup using custom segmentation software.
Results
In the uncemented cup, the mean BMD was 153 mg/cm3 with a between-scan difference of 10 mg/cm3 (P < 0.0001). In the cemented cup, the mean BMD was 186 mg/cm3 with a between-scan difference of 6 mg/cm3 (P = 0.15). In both uncemented and cemented cups the intraclass correlation coefficient between repeated measurements was >0.95 and narrow Bland–Altman Limits of Agreement.
Conclusion
BMD can be measured with high absolute between-scan agreement and good reliability adjacent to acetabular cemented and uncemented cups using DECT and segmentation software.
Keywords :
Segmentation , computed tomography bone density , dual energy computed tomography , bone loss , hip loosening cup