Author/Authors :
Mohsenzadeh, B. Biomedical Engineering and Medical Physics Department - Faculty of Medicine - Shahid Beheshti University of Medical Sciences, Tehran, Iran , Deevband, M.R. Biomedical Engineering and Medical Physics Department - Faculty of Medicine - Shahid Beheshti University of Medical Sciences, Tehran, Iran , Paydar, R. Department of Radiology - Iran University of Medical Sciences and Health Services, Tehran, Iran , Ghorbani, M. Biomedical Engineering and Medical Physics Department - Faculty of Medicine - Shahid Beheshti University of Medical Sciences, Tehran, Iran
Abstract :
Background: The present study aimed to investigate patient dose in common X-ray examinations to estimate effective dose in the digital radiography in Iran. Materials and Methods: Entrance surface dose (ESD) was measured based on applied exposure parameters for the common actual examination; and then effective dose (ED) was calculated by use of PCXMC software. The study was conducted on 15358 patients in 85 X-ray rooms; and the necessary data was collected for five age groups, 0-1 year, 1-5 years, 5-10 years, 10-15 years old and adults in each projection. Results: The ranges of ESD and ED in different examinations for all the age groups are 0.02-10.20 mGy and 2.42-378.96 µSv respectively. Conclusion: The effective dose as criteria can be used to reduce patients' doses. The special considerations such as: adequate training of imaging staff; updating clinical audits; patient dose considerations; implementation of systematic and regular quality assurance and quality control programs in medical imaging departments should be taken into account to optimize radiological practices.
Keywords :
Patient dose , dose reference level , dosimetry in diagnostic radiology , entrance surface dose , effective dose