Abstract :
Co(II) ion was determined by a new, accurate, sensitive and rapid method via a continuous flow injection analysis (CFIA) with a chemiluminescence reaction based on the oxidation of Luminol which is loaded on poly acrylic acid gel beads by hydrogen peroxide in presence of Cobalt (II) ion as a chemiluminescence catalyst. Chemical and physical parameters were investigated to obtain the best conditions. Linear dynamic range of Cobalt (II) ion was from 0.1-20.0 μg.ml^-1 with a correlation coefficient r = 0.9758, limit of detection (L.O.D) 0.2 ng/sample from the step wise dilution of lowest concentration in the calibration graph with the percentage relative standard deviation for 3 μg.ml-1 Co(ll) solution is 0.8537% (n=5). The method was applied successfully for the determination of Co (II) ion in CoMo catalyst type ( K F124- 3E).
Keywords :
Chemiluminescence , Flow injection analysis , gel beads , Cobalt (ll) Ion. , CoMo Catalyst (K F124 , 3E).