Title of article :
Preparation Pilocarpine Hydrochloride Selective Electrodes
Author/Authors :
abass, amina m. al-nahrain university - college of science - department of chemistry, Iraq
From page :
13
To page :
19
Abstract :
New eight selective electrodes of pilocarpine hydrochloride deepened on complexes of (PCH) -molybdophosphoric acid and (PCH) – phosphotungstate were prepared with varied plasticizers. Electrode with complex (DBPH-MP) was used as an active material, gave linear range from 5.0×10^-2 to 6.0×10^-5 M, and slope was 62.01 mV/decade, with detection limit was 2.0×10-5M, lifetime was near to 60 days. Electrode,s membrane made with active material (DBP-MP), showed concentration range was 2.1×10^-3-6.3×10^-5M, the slope was near to 52.34 mV/decade. Detection limit was 1.0×10^-5M, 55 days was the life time of this electrode. The parameters of electrode deepened on (DBPH-PT), concentration range was from 5.0×10^-2 to 6.5×10^-5 M with slope 50.41 mV/decade, 2.2×10^-5 M was the detection limit.Life time was determined around to 52 days.
Keywords :
Pilocarpine reactions , Sensors , membrane of PVC
Journal title :
Al-Nahrain Journal Of Science
Journal title :
Al-Nahrain Journal Of Science
Record number :
2647159
Link To Document :