• Title of article

    QUANTITATION OF ALFUZOSIN HYDROCHLORIDE IN PHARMACEUTICAL FORMULATIONS BY RP-HPLC

  • Author/Authors

    GANESH, MANI Nandha College of Pharmacy, India , UPPATYAY, SATISH Nandha College of Pharmacy, India , TIVARI, RISHI Nandha College of Pharmacy, India , KAMALAKANNAN, KALIAPPAN Nandha College of Pharmacy, India , RATHINAVEL, GOVINDASAMY Nandha College of Pharmacy, India , GANGULLY, SWASTIKA Biral Institute of Technology - Department of Pharmaceutical Sciences, India , SIVAKUMAR, THANGAVEL Nandha College of Pharmacy, India

  • From page
    263
  • To page
    266
  • Abstract
    An isocratic reversed phase high-performance liquid chromatographic (HPLC) method with ultraviolet detection at 245 nm has been developed for the determination of alfuzosin hydrochloride in dosage formulation. Good chromatographic separation alfuzosin was achieved by using a stainless steel analytical column Inertsil ODS-3V (5μm, 15 cm x 0.46 cm). The system was operated at ambient temperature (25 ±2°C) using a mobile phase consisting of acetonitrile : water : tetrahydrofuran : perchloricacid (250:740:10:1) at a flow rate of 1 ml/min. The calibration curve for alfuzosin hydrochloride was linear over the tested concentration range of 50%, 75%, 100%, 125% and 150% with reference to the label claim and a correlation coefficient of 0.999. The intra- and inter-run precision and accuracy results were 98.07 to 100.34 with the %RSD of 0.71% and tailings factor 1.07. The proposed method was validated for its selectivity, linearity, accuracy, and precision. The method was found to be suitable for the quality control of alfuzosin hydrochloride in bulk drug as well as in formulation.
  • Keywords
    Alfuzosin , UV detection , RP , HPLC , dosage formulation , method validation
  • Journal title
    Pakistan Journal Of Pharmaceutical Sciences
  • Journal title
    Pakistan Journal Of Pharmaceutical Sciences
  • Record number

    2647231