Title of article :
The practical approach to discerning between atrial lead displacement and reversal in dual chamber pacemakers
Author/Authors :
Al Mehairi, M. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Hubaishi, Y. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Al Ghamdi, S. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Al Fagih, A. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia
Abstract :
We describe a 12-lead electrocardiogram image of a pacemaker malfunction due to atrial lead displacement. Facilitating the understanding of pacemaker timing cycles could explain the underlying mechanisms in this instance. This review utilizes the feature of the ventricular safety pacing interval to discern between atrial lead displacement and reversal.
Keywords :
Crosstalk , Ventricular safety pacing , Lead displacement , Pacemaker
Journal title :
journal of the saudi heart association
Journal title :
journal of the saudi heart association