    The practical approach to discerning between atrial lead displacement and reversal in dual chamber pacemakers

  • Author/Authors

    Al Mehairi, M. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Hubaishi, Y. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Al Ghamdi, S. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia , Al Fagih, A. Prince Sultan Cardiac Center - Department of Adult Cardiology, Saudi Arabia

  • From page
    103
  • To page
    105
  • Abstract
    We describe a 12-lead electrocardiogram image of a pacemaker malfunction due to atrial lead displacement. Facilitating the understanding of pacemaker timing cycles could explain the underlying mechanisms in this instance. This review utilizes the feature of the ventricular safety pacing interval to discern between atrial lead displacement and reversal.