Title of article :
Relative Thermoluminescence Response of TLD-100, TLD-200 and GE-doped Optical Fiber to 8.05 keV X-ray
Author/Authors :
Amini, Y. M. university of malaya - Faculty of Science - Radiation Physics Laboratory, Physics Department, Malaysia , AI- Ruqeishi, M. S. Mohammed university of malaya - Faculty of Science - Radiation Physics Laboratory, Physics Department, Malaysia , Chew, S. H. university of malaya - Faculty of Science - Plasma Research Laboratory, Physics Department, Malaysia , Wong, C. S. university of malaya - Faculty of Science - Plasma Research Laboratory, Physics Department, Malaysia
From page :
91
To page :
95
Abstract :
The responses of three types of thermoluminescence dosimeters (TLDs), namely the TLD-100, TLD-200 and Ge-doped optical fiber to soft X-ray with photon energy of 8.05 keV were measured. The X-ray source used is the vacuum spark plasma which is capable of producing intense coper Kα line radiation at energy of 8.05 keV. The output dosage of the vacuum spark X-ray source is measured by a calibrated ion chamber and it is found to increase with operating voltage. The TL sensitivities of TLD-100, TLD-200 and Ge-doped optical fiber are determined and results are normalized to their responses to 60Co gamma rays. The TL sensitivities are found to be 1.2 for TLD-100, 29.0 for TLD-200 and 13.8 for Ge-doped Optical Fiber.
Keywords :
TLD , vacuum spark , Soft X , ray
Journal title :
Malaysian Journal of Science
Journal title :
Malaysian Journal of Science
Record number :
2680316
Link To Document :
بازگشت