• Title of article

    COMPARISON BETWEEN BRANDED CBC VIALS VERSUS LABORATORY MADE VIALS

  • Author/Authors

    MUNIR, M. King Edward Medical University - Department of Pathology, Pakistan , HAMEED, AHMAD King Edward Medical University - Department of Pathology, Pakistan , WAQAS, NISHA King Edward Medical University - Department of Pathology, Pakistan , RAFIQ, HUMERA King Edward Medical University - Department of Pathology, Pakistan , HAMEED, AMBAREEN King Edward Medical University - Department of Pathology, Pakistan

  • From page
    263
  • To page
    268
  • Abstract
    Introduction: Disposable items like CBC vials make an integral part of clinical laboratories. The study was carried out to determine the difference in results of four different types of vials. Study Design: It was an observational comparative study. Materials and Methods: Four hundred blood samples were included in the study. CBC was carried out within 4 hrs on all the four types of vials i.e. branded high quality and low quality vials and lab made vials with standard washing and routine washing. Instrument used was Sysmex KX – 21. Result: All the parameters measured and compared showed no difference between four types of vials. P-value was not significant. Discussion: RBC count, WBC count, Platelet count, Haemoglobin, HCT, MCV, DLC percentage in all the four types of vials showed no statistical difference in the results. Conclusion: The Labs can make their own CBC vials. Washing practice in haematology department KEMU is of international laboratory standards.
  • Keywords
    CBC (complete blood count) , CBC vacutainers or vials , EDTA anticoagulant.