Title of article :
Evaluation of Annual Staff Doses and Radiation Shielding Efficiencies of Thyroid Shield and Lead Apron during Preparation and Administration of 131I, 81Kr, and 99mTc‐Labeled Radiopharmaceuticals
Author/Authors :
Khalili, Nayereh Department of Radiology and Radiotherapy - Allied Medical Sciences School - Tehran University of Medical Sciences,Tehran , Zakariaee, Salman Department of Medical Physics - Faculty of Paramedical Sciences - Ilam University of Medical Sciences, Ilam , Jazayeri Gharebaghi, Elahe Department of Radiology and Radiotherapy - Allied Medical Sciences School - Tehran University of Medical Sciences,Tehran , Salehi, Yalda Department of Nuclear Medicine - Vali‐Asr Hospital - Tehran University of Medical Sciences, Tehran, Iran , Changizi, Vahid Department of Radiology and Radiotherapy - Allied Medical Sciences School - Tehran University of Medical Sciences,Tehran
Abstract :
Nuclear medicine technicians would receive unavoidable exposures during the preparation and
administration of radiopharmaceuticals. Based on the staff dose monitoring, the dose reduction
efficiencies of the radiation protection shields and the need to implement additional strategies to
reduce the staff doses could be evaluated. In this study, medical staff doses during the preparation
and administration of Tc‐99 m, I‐131, and Kr‐81 radiopharmaceuticals were evaluated. The
dose reduction efficiencies of the lead apron and thyroid shield were also investigated. GR‐207
thermoluminescent dosimeter (TLD) chips were used for quantifying the medical staff doses. The
occupational dose magnitudes were determined in five organs at risk including eye lens, thyroid,
fingers, chest, and gonads. TLDs were located under and over the protective shields for evaluating
the dose reduction efficiencies of the lead apron and thyroid shield. The occupational doses were
normalized to the activities used in the working shifts. During preparation and injection of Tc‐99 m
radiopharmaceutical, the average annual doses were higher in the chest (4.49 mGy) and eye lenses
(4 mGy). For I‐131 radiopharmaceutical, the average annual doses of the point‐finger (15.8 mGy)
and eye lenses (1.23 mGy) were significantly higher than other organs. During the preparation and
administration of Kr‐81, the average annual doses of the point‐finger (0.65 mGy) and chest (0.44
mGy) were higher. The significant dose reductions were achieved using the lead apron and thyroid
shield. The radiation protection shields and minimum contact with the radioactive sources, including
patients, are recommended to reduce the staff doses.
Keywords :
Annual staff doses , lead apron , nuclear medicine , occupational dose , thermoluminescent dosimeter , thyroid shield
Journal title :
Journal of Medical Signals and Sensors (JMSS)