Title of article :
Label-Free Perfect Separation of Michigan Cancer Foundation-7 Cells from White Blood Cells using Both Wall-Induced Lift and Dielectrophoresis Forces
Author/Authors :
Alaghemand ، Alireza Electrical Engineering Department - Shahrood University of Technology , Fattah ، Ali Electrical Engineering Department - Shahrood University of Technology
Abstract :
A microchannel device was designed and utilized to separate Michigan Cancer Foundation-7 cells from white blood cells using both wall-induced lift and di-electrophoresis forces. Using COMSOL Multiphysics simulator, the label-free separation process is performed based on the radius size of the particles. The best performance of the structure is obtained with the efficiency of 100% and 99% for cancer and white blood cells separation processes, respectively. For the proposed microfluidic structure, in addition to the highest available efficiency and high separation speed of 12 um/s, the low DC voltage of 6 V is applied, which causes negligible damage to the blood cells.
Keywords :
Cancer cells , COMSOL , Dielectrophoresis , Microfluidics , separation
Journal title :
International Journal of Advanced Design and Manufacturing Technology
Journal title :
International Journal of Advanced Design and Manufacturing Technology