Title of article :
An Apparatus for Specific Heat Capacity Measurement by Thermal Radiation Calorimetry
Author/Authors :
K. Morimoto، نويسنده , , S. Sawai and K. Hisano ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
11
From page :
709
To page :
719
Abstract :
Specific heat capacity measurements of disk-shaped specimens have been performed with an apparatus based on thermal radiation calorimetry. The specimen surfaces were irradiated by two fiat heaters in a vacuum chamber so that homogeneous temperature distribution within an insulating specimen was achieved. Homogeneity was confirmed by computer simulation based on the controlvolume method. The values of specific heat capacity were obtained by measurement of the specimen temperature, the time rate change of the specimen temperature, and the radiant power from the heater for heating and cooling modes. The specific heat capacities of Ni metal, A12O3 ceramic, MgO ceramic, and A1N ceramic were measured in the temperature range from 220 to 500oC to confirm the validity of this calorimeter. The relative error involved in the measured values was estimated to be ±3%.
Keywords :
Emissivity , Calorimetry , high temperature specific heat capacity , thermal radiation.
Journal title :
International Journal of Thermophysics
Serial Year :
1999
Journal title :
International Journal of Thermophysics
Record number :
426509
Link To Document :
بازگشت