• Title of article

    An Apparatus for Specific Heat Capacity Measurement by Thermal Radiation Calorimetry

  • Author/Authors

    K. Morimoto، نويسنده , , S. Sawai and K. Hisano ، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    11
  • From page
    709
  • To page
    719
  • Abstract
    Specific heat capacity measurements of disk-shaped specimens have been performed with an apparatus based on thermal radiation calorimetry. The specimen surfaces were irradiated by two fiat heaters in a vacuum chamber so that homogeneous temperature distribution within an insulating specimen was achieved. Homogeneity was confirmed by computer simulation based on the controlvolume method. The values of specific heat capacity were obtained by measurement of the specimen temperature, the time rate change of the specimen temperature, and the radiant power from the heater for heating and cooling modes. The specific heat capacities of Ni metal, A12O3 ceramic, MgO ceramic, and A1N ceramic were measured in the temperature range from 220 to 500oC to confirm the validity of this calorimeter. The relative error involved in the measured values was estimated to be ±3%.
  • Keywords
    Emissivity , Calorimetry , high temperature specific heat capacity , thermal radiation.
  • Journal title
    International Journal of Thermophysics
  • Serial Year
    1999
  • Journal title
    International Journal of Thermophysics
  • Record number

    426509