Title of article :
Thermal Diffusivity Measurements of Thermographite
Author/Authors :
K.D. Magli? and N.D. Milo?evi? ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
This paper presents results of measurements of a graphite proposed to serve as a
thermophysical property reference or standard reference material. The reported
measurements contribute to a program launched in 1999 by AInter Corp. with
the objective to provide a replacement for the NIST thermal property reference
material RM AXM-5Q graphite whose supplies were being exhausted. Measurements
of the thermal diffusivity performed on five specimens taken from
different positions within a large graphite block between room temperature and
1300 K were in good mutual agreement. Measurements of NIST reference AXM-
5Q graphite sample supplied to minimize effects of different contributors to a
common base were also in good agreement, both with the NBS reference function
established by Hust in 1984 and contributions to the NBS project from the
Vincˇa Institute of Nuclear Sciences carried out in 1979. The influence of different
data reduction techniques on the measured thermal diffusivity values is
illustrated and discussed.
Keywords :
AXM-5Q graphite , standard reference materials , thermal diffusivity , thermographite , thermophysical property SRMs.
Journal title :
International Journal of Thermophysics
Journal title :
International Journal of Thermophysics