Title of article :
Overcoming Inhomogeneity and Hysteresis Limitations of Type R Thermocouples in an IInternational Comparison
Author/Authors :
Ferdouse Jahan and Mark Ballico ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Type R thermocouples are widely used and convenient high-temperature
transfer standards; however, the achievable accuracy is limited by the
effects of inhomogeneity and hysteresis. In this article, we summarize the results
of the recent iInternational comparison APMP-T-S1-04 and discuss the results of the
thermoelectric scanning, spatially resolved over the length of the thermocouples. The
thermoelectric signatures show both reversible (hysteresis) and irreversible inhomogeneities
introduced by the calibration processes used by the participants. The results
demonstrate that although the reversible hysteresis of Type R thermocouples limits
their performance as a transfer standard in thermometry, this can be managed by
appropriate design of the comparison protocol. By performing all calibrations from
lower to higher temperatures from an initial 450◦C anneal state, a pilot laboratory
reproducibility of typically 0.03◦C (k = 2) and a reference value uncertainty of 0.03–
0.06◦C (at k = 2) over 0–1,100◦C were achieved. This allowed statistically significant
testing of the calibration capabilities of all the participants.
Keywords :
Annealing state · Inhomogeneity · IIntercomparison · Stability ·Type R thermocouple
Journal title :
International Journal of Thermophysics
Journal title :
International Journal of Thermophysics