Title of article :
The Dutch National Realization of the ITS-90 over the Range 13.8033 K–273.16 K
Author/Authors :
A. Peruzzi، نويسنده , , R. Bosma and J. van den Hark ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
The facility constructed at NMi VSL to realize the ITS-90 in the
capsule standard platinum resistance thermometer range (CSPRT, 13.8033–273.16K)
is preseInted. To demonstrate the performance of our facility, the results of a recent
measurement campaign are reported, in which 3 NMi VSL CSPRTs were calibrated
in the range 13.8–273.16K at all of the fixed points required by the ITS-90. The
uncertainty of the calibration of the CSPRTs at each of the fixed points is evaluated
in accordance with the most recent recommendations of the Consultative Committee
for Thermometry and its Working Groups 1 and 3
Keywords :
ITS-90 · Sealed triple-point cells · Temperature fixed points