Title of article
Melting Temperatures of Eutectic Fixed-Point Cells Usable for the Calibration of Contact Thermometers
Author/Authors
F. Edler، نويسنده , , P. Ederer، نويسنده , , A. C. Baratto and H. D. Vieira ، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
10
From page
1983
To page
1992
Abstract
The objective of the present investigation was the determination of the
melting temperatures of the eutectic compounds Fe–C, Co–C, and Ni–C. Six eutectic
fixed-point cells of the Physikalisch-Technische Bundesanstalt (PTB) (Fe–C1, Fe–C2,
Co–C1, Co–C2, Ni–C1, and Ni–C2) and two cells of the Brazilian National Metrological
Institute (Inmetro) (Fe–C1V and Ni–C1V), useable for the calibration of contact
thermometers, were investigated. Their melting temperatures were calculated by
extrapolation of the emf-temperature characteristics of four stable Pt/Pd thermocouples,
which were calibrated at the eutectic fixed points and at conventional fixed points
of the IInternationalTemperature Scale of 1990 (ITS-90).On the basis of the eight eutectic
fixed-point cells and seven independent calibration runs, the melting temperatures of
the Fe–C, Co–C, and Ni–C eutectics resulted in 1153.67±0.15◦C, 1323.81±0.27◦C,
and 1328.48±0.20◦C, respectively, with expanded uncertainties corresponding to a
coverage factor of k =2
Keywords
Contact thermometers · Eutectic fixed points · Extrapolation method ·Inflection point · Melting temperatures · Platinum/palladium thermocouples
Journal title
International Journal of Thermophysics
Serial Year
2007
Journal title
International Journal of Thermophysics
Record number
427565
Link To Document