Title of article :
Response of the Active Scattering Aerosol Spectrometer Probe (ASASP-100X) to Particles of Different Chemical Composition
Author/Authors :
Kim، نويسنده , , Young J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
10
From page :
33
To page :
42
Abstract :
Response of the Active Scattering Aerosol Spectrometer Probe (ASASP-I00X) was simulated numerically for particles of different chemical composition. Simulation was based on the Mie scattering calculations of the ASASPʹs response to input lognormal size distributions of particles having different refractive indices. Results from such simulation calculations show that the measurement data would greatly distort the true size distri-butions with apparent artificial peaks and underestimation of particle volume, especially in the resonant region of the instrumentʹs response, 0.4-0.8 /-tm particle diameter size range. Thus it is very important in interpreting the size distribution data measured with an ASASP to correct them using the size calibration data based on the optical properties of the particles being measured.
Journal title :
Aerosol Science and Technology
Serial Year :
1995
Journal title :
Aerosol Science and Technology
Record number :
430219
Link To Document :
بازگشت