Title of article :
The Particle Detection Efficiency Curve of the TSI-3010 CPC as a Function of the Temperature Difference between Saturator and Condenser
Author/Authors :
Mertes، نويسنده , , S.; Schr?der، نويسنده , , F.; Wiedensohler، نويسنده , , A، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
Particle detection efficiences of the Condensation Particle
Counter (CPC) TSI-3010 are measured for various
temperature differences between saturator and
condenser. By increasing the temperature difference
from the default value of 17°C to the maximum value of
2SoC, the particle detection efficiency curve becomes
steeper and shifts to smaller particle diameters. In this way, the50% detection efficiency diameter of the TSI·
3010 CPC decreases from 10.S to S.7 nm, without the
necessity to change its internal design. In this size
range, the desired lower detection limit of the TSI-3010
can be adjusted by operating the CPC at the corresponding
temperature difference between saturator and
condenser.
Journal title :
Aerosol Science and Technology
Journal title :
Aerosol Science and Technology