Abstract :
Cascade impactor analysis of drugs for inhalation in aerosol form is
regarded by both European and U.S. pharmacopoeias as an important method to
determine particle size-related information concerning dose that might be delivered
to various parts of the respiratory tract. Existing measurements usually are made at
flow rates in excess of 25 l/min, which are more representative of peak inspiratory
flow rates achieved by adults. There is therefore a need for a low-flow impactor that
can be used to evaluate drug delivery devices intended for use by infants or small
children, or by those whose disease status prevents the achievement of higher
inspiratory flow rates. A new version of the Marple-Miller impactor™ (MMI™) is
described that meets this need, operating at either 4.9 or 12 l/min by the substitution
of a different uppermost stage jet. A novel feature is the use of external stage
collection cups instead of plates, enabling quantitative transfer of drug collecting on
each stage for subsequent chemical assay. Calibration data and an assessment of
inter-stage losses are presented for each of the flow rates
Journal title :