Title of article :
Reliability systems for implantable cardiac defibrillator batteries
Author/Authors :
Esther S. Takeuchi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
5
From page :
115
To page :
119
Abstract :
The reliability of the power sources used in implantable cardiac defibrillators is critical due to the life-saving nature of the device. Achieving a high reliability power source depends on several systems functioning together. Appropriate cell design is the first step in assuring a reliable product. Qualification of critical components and of the cells using those components is done prior to their designation as implantable grade. Product consistency is assured by control of manufacturing practices and verified by sampling plans using both accelerated and real-time testing. Results to date show that lithium/silver vanadium oxide cells used for implantable cardiac defibrillators have a calculated maximum random failure rate of 0.005% per test month.
Keywords :
Reliability , Implantable medical applications