• Title of article

    Lithium-ion batteries for hearing aid applications: II. Pulse discharge and safety tests

  • Author/Authors

    S. Passerini، نويسنده , , F. Coustier، نويسنده , , B. B. Owens، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    9
  • From page
    144
  • To page
    152
  • Abstract
    Rechargeable lithium-ion batteries were designed to meet the power requirements of hearing aid devices (HADs). The batteries were designed in a 312-button cell size, compatible with existing hearing aids. The batteries were tested to evaluate the design and the electrochemical performance, as they relate to a typical hearing aid application. The present report covers the pulse capabilities, cycle life and preliminary safety tests. The results are compared with other battery chemistries: secondary lithium-alloy and nickel–metal hydride batteries and primary Zn–air batteries. The cell AC impedance was stable over the frequency range between 1 and 50 kHz, ranging between 5 Ω at the higher frequency and 12 Ω at the lower extreme. Pulse tests were consistent with these values, as the cells were capable of providing a series of 100 mA pulses of 10-s duration. The safety tests suggest that the design is intrinsically safe with respect to the most common types of abuse conditions.
  • Keywords
    safety , Hearing aid cell , Li-ion button cell , Hearing aid
  • Journal title
    Journal of Power Sources
  • Serial Year
    2000
  • Journal title
    Journal of Power Sources
  • Record number